SECTION 2—ASSEMBLY

SECTION 2—ASSEMBLY

Assembling the Patient Lift

WARNING

Tighten mounting screws securely. Failure to do so may result in bodily injury or damage to the lift.

NOTE: For this procedure, refer to FIGURE 2.1 on page 12.

NOTE: The Reliant 440 is packed partially assembled for shipping purposes.

1.Remove all contents from the box and place the unit on a workbench or table.

2.Remove the two mounting screws, washers, and locknuts connecting the mast to the base (Detail “A”).

3.Remove the mast from the base.

4.Position and re‐attach the mast to the base using the two mounting screws, washers, and locknuts removed in STEP 2 (Detail “B”). Securely tighten.

5.Insert the pins of the rod ends up through the tabs on the base (Detail “B”).

6.Attach the pins of the rod ends to the tabs using the two locknuts and washers. Securely tighten.

7.Cut the cable tie securing the actuator (Detail “C”).

8.Swing the actuator forward so that it is out of the way.

9.Swing the boom up and over to the front.

10.Attach the push handle to the back of the mast using the two mounting screws, locknuts, and washers which are provided (Detail “D”).

11.Attach the kneepad. Refer to Attaching the Kneepad on page 13.