SECTION 7— TROUBLESHOOTING
SYMPTOMS
FAULTS
SOLUTION
Mast/base joint loose. | Tighten the mounting screws that secure the | |
|
| mast to the base. Refer to Assembling the Patient |
|
| Lift on page 11. |
| Tie - rods are loose. | Refer to Maintaining the Base Adjustment on |
|
| page 34. |
|
|
|
Casters/brakes noisy or | Fluff or debris in bearings. | Replace casters. |
stiff. |
|
|
|
|
|
Electric actuator fails to | Check connections. | |
lift when button is | nector loose. |
|
pressed. |
|
|
| Batteries low. | Charge batteries. Refer to Charging the Battery |
|
| on page 19. |
| RED emergency stop button | Rotate RED emergency stop button CLOCK- |
| pressed IN. | WISE until it pops out. |
| Battery not connected prop- | Reconnect the battery to the control box. Refer |
| erly to control box. | to Charging the Battery on page 19. |
| The connecting terminals are | Replace the battery pack. Refer to Charging the |
| damaged. | Battery on page 19. |
| Electric actuator in need of ser- | Refer to Replacing Electric Actuator on page 33. |
| vice or load is too high. | Contact your Dealer. |
|
|
|
Unusual noise from | Actuator is worn or damaged | Refer to Replacing Electric Actuator on page 33. |
actuator. | or spindle is bent. | Contact your Dealer. |
|
|
|
Lift arms will not lower | Lift arms require a minimum | Pull down slightly on the lift arms. |
in uppermost position. | weight load to lower from the |
|
| uppermost position. |
|
|
|
|
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.