SECTION 7—TROUBLESHOOTING

SECTION 7— TROUBLESHOOTING

SYMPTOMS

FAULTS

SOLUTION

Stand-up lift feels loose.

Mast/base joint loose.

Tighten the mounting screws that secure the

 

 

mast to the base. Refer to Assembling the Patient

 

 

Lift on page 11.

 

Tie - rods are loose.

Refer to Maintaining the Base Adjustment on

 

 

page 34.

 

 

 

Casters/brakes noisy or

Fluff or debris in bearings.

Replace casters.

stiff.

 

 

 

 

 

Electric actuator fails to

Hand-control or actuator con-

Check connections.

lift when button is

nector loose.

 

pressed.

 

 

 

Batteries low.

Charge batteries. Refer to Charging the Battery

 

 

on page 19.

 

RED emergency stop button

Rotate RED emergency stop button CLOCK-

 

pressed IN.

WISE until it pops out.

 

Battery not connected prop-

Reconnect the battery to the control box. Refer

 

erly to control box.

to Charging the Battery on page 19.

 

The connecting terminals are

Replace the battery pack. Refer to Charging the

 

damaged.

Battery on page 19.

 

Electric actuator in need of ser-

Refer to Replacing Electric Actuator on page 33.

 

vice or load is too high.

Contact your Dealer.

 

 

 

Unusual noise from

Actuator is worn or damaged

Refer to Replacing Electric Actuator on page 33.

actuator.

or spindle is bent.

Contact your Dealer.

 

 

 

Lift arms will not lower

Lift arms require a minimum

Pull down slightly on the lift arms.

in uppermost position.

weight load to lower from the

 

 

uppermost position.

 

 

 

 

NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.