SECTION
|
|
| Side View |
|
Rear of Chair |
| Seat Cushion | Front of Chair | |
|
|
| ||
|
|
|
| Side Frame |
|
|
|
| Footrest |
DO NOT Remove |
|
|
|
|
This Bolt | Bolt A | Bolt B | Bolt C Bolt D | Recline Mechanism |
| Guard | |||
|
|
|
|
FIGURE 5.4 Replacing the Recline Mechanism Guard
Replacing the Elbow Pads - Model IH6077A Only
NOTE: For this procedure, refer to FIGURE 5.5.
1.Remove two hex head screws that secure the existing elbow pad to the elbow pad bracket.
2.Remove existing elbow pad and discard.
3.Using the existing hex head screws, secure the new elbow pad to the elbow pad bracket.
Adjusting Gas Spring Recline Cylinder Assembly - Model IH6077A Only
NOTE: For this procedure, refer to FIGURE 5.6 on page 25.
1.Remove the mounting screw which secures the cable to the back assembly.
2.Remove the retaining ring from the clevis pin.
3.Remove mounting bolt and locknut securing the gas spring recline cylinder to the back frame.
4.Recline the back until cylinder can be removed from mounting bracket easily (approximately a 45° angle).
5.Remove gas spring recline cylinder.
6.Loosen the locknut on the bottom of the cylinder (Detail ʺAʺ).
7.Hold the bottom bracket and turn the cylinder until cable becomes taut.
NOTE: If the cylinder was not locking when the lever was actuated, turn cylinder clockwise. If the cylinder was not locking when the lever was released, turn cylinder counterclockwise.
Invacare Mobile Recliners | 24 | Part No. 1141494 |