SECTION 3—INSTALLING THE CONSERVING REGULATOR

NOTE: For information on how to operate the cylinder, consult your equipment provider.

NOTE: For this procedure, refer to FIGURE 3.1.

1.Loosen the cylinder post valve T‐Handle.

2.Lower the conserver over any CGA870 post valve cylinder with the alignment pins toward the holes on the cylinder neck.

3.Align the two pins and gasket with the

corresponding holes on

FIGURE 3.1 Attaching the

the cylinder post valve.

IOC100P to the Cylinder

NOTE: Cylinder post valve T‐Handle should be aligned with indentation on post valve.

4.While holding the unit in place, tighten the cylinder post valve by turning clockwise.

5.Attach a standard cannula [7 ft. (2.13 m) or less in length] to the oxygen supply outlet fitting.

NOTE: Tighten only by hand. The use of a tool to tighten the T‐Handle may damage the unit.