SECTION 4—TYPICAL PRODUCT PARAMETERS

Flow Range:

Platinum 5 and XL - 0.5 to 5 L/min For

 

flowrates less than 1 L/min, we recommend

 

the use of the Invacare Pediatric Flowmeter

 

IRCPF16 Accessory.

 

 

 

Platinum 10 - 2 L/min to 10 L/min

 

 

Low Flow Alarm:

Platinum 5 - 0 L/min to 0.5 L/min,

 

Platinum 10 - 0 L/min to 1 L/min

 

Rapid Audible Alarm Beeping (No

 

Accessories Connected)

 

 

Average Power

Platinum 5 and XL: 400W

Consumption:

Platinum 10: 585W

 

 

Pressure Relief Mechanism

35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)

Operational at:

 

 

 

Change in maximum recom-

0.7 L/min

mended flow when back pres-

 

sure of 7kPa is applied:

 

 

 

Filters:

Cabinet (2), Outlet HEPA and Compressor

 

Inlet.

 

 

Safety System:

Current overload or line surge shutdown.

 

High temperature compressor shutdown.

 

High Pressure Alarm w/compressor shut-

 

down. Low Pressure Alarm w/compressor

 

shutdown. Battery Free Power Loss Alarm.

 

SensO2 Oxygen System (SensO2 Model)

 

Low Flow Alarm.

 

 

Width:

18 3/8 in ± 3/8 in (46.7 cm ± 1 cm)

 

 

Height:

26 3/8 in ± 3/8 in (67.0 cm ± 1 cm)

 

 

Depth:

14 3/8 in ± 3/8 in (36.5 cm ± 1 cm)

 

 

Weight:

Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ±

 

1 kg)

 

 

 

Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg)

 

 

Shipping Weight:

Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ±

 

1 kg)

 

 

 

Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg)

 

 