SECTION 4—TYPICAL PRODUCT PARAMETERS

SECTION 4—TYPICAL PRODUCT PARAMETERS

Electrical Requirements:

120 VAC ± 10% (132 VAC/108 VAC), 60

 

Hz

 

 

Rated Current Input:

Platinum 5 and XL: 4.3 A

 

 

 

Platinum 10: 4.7 A

 

 

Sound Level:

Platinum 5 and XL: 50 dB Average

 

 

 

Platinum 10: 58 dB Average

 

 

Altitude:

Platinum 5 and XL - Up to 6,000 ft (1828

 

meters) above sea level without degradation

 

of concentration levels. From 6,000 ft (1828

 

meters) to 13,129 ft (4000 meters) below

 

90% efficiency

 

 

 

Platinum 10 - Up to 4,000 ft (1230 meters)

 

above sea level without degradation of con-

 

centration levels. From 4,000 ft (1230

 

meters) to 13,129 ft (4000 meters) below

 

90% efficiency

 

 

*Oxygen Output

All 5LXO2/5LX models 87% to 95.6% for

Concentration Levels:

flows ranging 0.5 to 5 L/min and all

 

10LXO2/10LX models 87% to 95.6% for

 

flows ranging 2 L to 10 L/min

 

 

Maximum Outlet Pressure:

Platinum 5 and XL: 5 psi ± 0.5 psi (34.5 kPa

 

± 3.45 kPa)

 

 

 

Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45

 

kPa)

*NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).