SECTION 5—OPERATING INSTRUCTIONS

Front Panel Indicators - Units with SensO2

NOTE: For this procedure, refer to FIGURE 5.7.

The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.

OXYGEN PURITY

Auto Shut - Down

Use Backup

RED

Call Supplier

 

Below Normal

YELLOW

Normal O2

GREEN

FIGURE 5.7 Front Panel Indicators - Units with SensO2

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.

LABEL

 

O2 PURITY

INDICATOR LIGHTS

SYMBOL

 

 

(LED)

 

 

 

O2

SYSTEM OKAY

GREEN Indicator Light

O2 over 85% (± 2%)

 

 

O Between 73% (± 3%)

YELLOW Indicator light

 

2

 

A. YELLOW Solid

 

 

 

 

to 85% (±2%)

B. YELLOW Flashing Sensor Failure

 

 

 

Call a qualified technician. Have a

 

 

 

back-up supply of oxygen ready.

 

 

 

 

SYSTEM FAILURE

RED Indicator Light

 

O2 Below 73% (±3%)

Continuous Audible Alarm

 

 

 

Sieve-GARD Compressor Shut-

 

 

 

down. Call a qualified technician.

 

 

 

Switch to a back-up supply of oxy-

 

 

 

gen immediately.

Part No 1118353

21

PlatinumSeries

Page 21
Image 21
Invacare IRC5LX, IRC10LXO2 manual Front Panel Indicators Units with SensO2, Initial Startup of the Concentrator