SECTION 7—TROUBLESHOOTING GUIDE
|
|
|
|
SYMPTOM: | PROBABLE CAUSE: |
| SOLUTION: |
Alarm: | Main Power Loss: | 1. | Insert plug into outlet. |
SHORT BEEPS, | 1. Power cord NOT |
|
|
LONG PAUSE | plugged in. |
|
|
|
|
|
|
Concentrator | 2. No power at outlet. | 2. | Inspect house circuit break- |
| ers or fuses. If problem recurs, | ||
NOT operating, |
| use a different outlet. | |
power switch ON. |
|
|
|
3. Tripped circuit breaker. | 3. | Call Service Provider. | |
Beep......Beep |
|
|
|
|
|
| |
Alarm: | System Failure: | 1a. Remove and clean cabinet | |
CONTINUOUS | 1. Unit overheating due to | filters. | |
| blocked air intake. |
| |
Concentrator | 1b. Move oxygen concentrator | ||
| at least three inches away from | ||
NOT operating, |
| walls, draperies or furniture. | |
Power Switch ON. |
|
|
|
2. Insufficient power at | 2. | DO NOT use extension | |
Beeeeeeeep.... | outlet. | cords. Move to another elec- | |
| trical outlet or circuit. | ||
|
| ||
|
|
|
|
| 3. Internal repairs | 3. | Call Service Provider. |
| required |
|
|
|
|
|
|
Alarm: | Low Flow Alarm: | 1. | Inspect for kinks or |
RAPID | 1. Kinked or blocked tub- | blockages. Correct, clean or | |
| ing, cannula or humidifier. | replace item. Once | |
Beep....Beep... |
| corrected, turn power OFF | |
Beep...Beep |
| for 60 seconds and then turn | |
|
| power back ON. | |
|
|
|
|
| 2. Flowmeter set too low: | 2. | Reset flowmeter to pre- |
| 0.5 L/min (or less) for | scribed flowrate. If prescribed | |
| Platinum 5; 1.0 L/min (or | flowrate is below 0.5 L/min, | |
| less) for Platinum 10. | use the Pediatric Flowmeter | |
|
| IRCPF16 (Only with Platinum | |
|
| 5 and XL). | |
|
|
|
|