SECTION 7—TROUBLESHOOTING GUIDE

 

 

 

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

YELLOW or RED

1. Low oxygen purity.*

1. Clean or Replace filters.

LIGHT Illuminated.

 

 

2. Kinked or blocked tub-

2. Inspect for kinks or block-

Alarm:

ing, cannula or humidifier.*

ages. Correct, clean or replace

 

item. Once corrected, turn

CONTINUOUS

 

power OFF for 60 seconds and

On red light only.

 

then turn power back ON.

 

 

 

* Only applies to

3. Flowmeter set too low:

3. Check flowmeter is set to

0.5 L/min (or less) for

1.0 L/min. or more.

5LXO2 and

Platinum 5; 1.0 L/min (or

 

10LXO2 SensO2

less) for Platinum 10.

 

models.

 

 

4. Unit overheating due to

4.a. Remove and clean cabinet

 

blocked air intake.

filters.

 

 

 

 

 

4b. Move concentrator at least

 

 

three inches from walls, drap-

 

 

eries, and furniture.

 

 

 

 

5. Insufficient power at

5. DO NOT use extension

 

outlet.

cords. Move to another elec-

 

 

trical outlet or circuit.

 

 

 

 

6. Internal repairs

6. Call Service Provider.

 

required.

 

 

 

 

GREEN LIGHT

1. Internal repairs

1. Call Service Provider.

with YELLOW

required.

 

Light Flashing.

 

 

 

 

 

Part No 1118353

27

PlatinumSeries

Page 27
Image 27
Invacare IRC5LX, IRC10LXO2 manual Troubleshooting Guide Symptom