SECTION
On/Off (I/O)
Power Switch
FIGURE 5.5 Power Switch
Flowrate
NOTE: For this procedure, refer to FIGURE 5.6 on page 25.
1.Turn the flowrate knob to the setting prescribed by your physician or therapist.
WARNING
DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed.
2.If the flowrate on the flowmeter ever falls below 0.5 L/min. for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than 0.5 L/min., the LOW FLOW audible alarm will go off.
Platinum™ | 24 | Part No 1134866 |