Invacare IRC5LXAW, IRC5LXO2AWQ, IRC9LXO2AWQ manual

Models: IRC9LXO2AWQ IRC5LXAW IRC5LXO2AWQ IRC5LXO2AW

1 36
Download 36 pages 51.11 Kb
Page 15
Image 15

SECTION 4—TYPICAL PRODUCT PARAMETERS

*Oxygen Output

IRC5LXAW, IRC5LXO2AW and

Concentration Levels:

IRC5LXO2AWQ

All 5LXO2AWQ/5LXAW/

93% minimum at 1 to 3 L/min.

5LXO2AW/9LXO2AWQ

91% minimum at 4 L/min.

models

87% minimum at 5 L/min. (maximum

* (Stated concentration levels

recommended flow)

achieved after initial warm-up

IRC9LXO2AWQ

period (approximately 30

92% minimum at 1 to 2 L/min.

minutes).

91% minimum at 8 L/min.

 

87% minimum at 9 L/min.

 

93% minimum at 3 to 7 L/min.

 

 

Maximum Outlet Pressure:

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

34.5 ± 3.45 kPa (5 ± 0.5 psi)

 

 

 

IRC9LXO2AWQ

 

62.0 kPa ± 3.45 kPa (9 ± 0.5 psi)

 

 

Flow Range:

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

0.5 to 5 L/min. Flowrates less than 1 L/min.

 

are not recommend

 

 

 

IRC9LXO2AWQ

 

1 L/min. to 9 L/min. Flowrates less than 1

 

L/min. are not recommended.

 

 

Low Flow Alarm:

0 L/min. to 0.5 L/min. Rapid Audible Alarm

 

Beeping (No Accessories Connected).

 

 

Average Power Consumption:

(IRC5LXO2AWQ)

 

300 W

 

 

 

280 W @

>

3 L/min.

 

 

 

 

 

(IRC5LXO2AW)

 

365 W

 

 

 

340 W @

>

3 L/min.

 

 

 

 

 

380 W (IRC5LXAW)

 

 

 

475 W (IRC9LXO2AWQ)

 

 

Pressure Relief Mechanism

241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)

Operational at:

 

 

 

 

 

 

 

Part No 1134866

15

Platinum™

Page 15
Image 15
Invacare IRC5LXAW, IRC5LXO2AWQ, IRC9LXO2AWQ manual