Invacare IRC5LXO2AWQ, IRC5LXAW Recommended Guidelines for Optimum Performance, Plug in Power Cord

Models: IRC9LXO2AWQ IRC5LXAW IRC5LXO2AWQ IRC5LXO2AW

1 36
Download 36 pages 51.11 Kb
Page 20
Image 20

SECTION 5—OPERATING INSTRUCTIONS

Recommended Guidelines for Optimum Performance

Temperature:

(10°C - 35°C) 50°F - 95°F

 

 

Electrical:

No extension cords.

 

 

Placement:

No closer than 7.5 cm (3 in) from the wall,

 

furniture, draperies, or similar surfaces.

 

 

Tubing and Cannula:

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

2.1 m (7 ft) cannula with a maximum 15.2 m

 

(50 ft) of Crush-Proof Tubing (DO NOT

 

pinch).

 

 

 

IRC9LXO2AWQ

 

Recommended use up to 15.2 m (50 ft) high

 

flow tubing with high flow cannula at all flow

 

rates.

 

 

Environment:

Smoke and soot-free. No confined spaces

 

(Example: No closets).

 

 

Relative Humidity:

20 to 60%

 

 

Time of Operation:

Up to 24 hours per day.

 

 

Flow Rate

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

0.5 L/min. to 5 L/min. Flow rates less than 1

 

L/min., are not recommended.

 

 

 

IRC9LXO2AWQ

 

1 L/min. to 9L/min. Flow rates less than 1

 

L/min. are not recommended.

 

 

Minimum Operating Time:

30 Minutes

NOTE: Ensure that your concentrator is at least 7.5 cm (3 in) away from walls, draperies or furniture to assure sufficient air flow. Avoid deep pile carpets and heaters, radiators or hot air registers.

Plug in Power Cord

1. Plug in power cord to an electrical outlet.

Platinum™

20

Part No 1134866

Page 20
Image 20
Invacare IRC5LXO2AWQ, IRC9LXO2AWQ, IRC5LXAW manual Recommended Guidelines for Optimum Performance, Plug in Power Cord