SECTION 6—TROUBLESHOOTING
PROBLEM | SOLUTION |
|
|
Nebulizer overheats. The | Turn the unit OFF for a minimum of 40 |
nebulizer has an internal | minutes to allow the nebulizer to cool. |
thermal protector to |
|
prevent the device from |
|
overheating. If the internal |
|
temperature reaches the |
|
preset level, the unit will |
|
automatically shut down, and |
|
the power LED will turn |
|
RED. |
|
|
|
The nebulizer does not | Check all power sources to the nebulizer. |
operate. | Push the ON/OFF button again. |
|
|
Medication does not nebu- | Check for proper connection of the tubing |
lize properly. | adapter. |
| Make sure the nebulizer is in an upright position |
| and not tilted beyond 30º. |
| Make sure the nebulizer is not empty. |
| Clean the nebulizer as described in the manual. |
| Make sure the cone has been properly placed in |
| the nebulizer cup. |
| Replace the nebulizer. |
| Check the intake filter and replace it if it is dirty. |
|
|
Nebulizer delivers no mist. | Make sure the cone is in the cup and properly |
| seated. |
|
|
Stratos™Portable Plus | 32 | Part No 1150755 |