MODEL DO-03/04

SECTION 10.0

 

TROUBLESHOOTING

SECTION 10.0

TROUBLESHOOTING

10.1 OVERVIEW

The 54eA analyzer continuously monitors itself and the sensor for faults. When the analyzer detects a fault in the oxygen sensor or in the instrument itself it displays a fault message. If alarm 4 was enabled, the red FAIL LED will also light and relay 4 will activate. The outputs will go to 22.00 mA or to the value programmed in Section 5.5.

See Section 10.2 for an explanation of fault messages and suggested corrective actions.

The analyzer also displays error and warning messages if a calibration is seriously in error. Refer to the section below for assistance. Each section also contains hints for correcting other measurement and calibration problems.

 

Measurement

Section

 

 

Temperature

10.3

 

 

Dissolved oxygen

10.4

 

 

 

 

 

For troubleshooting not related to measurement problems, see Section 10.5.

To view diagnostic variables, go to the main display and press any key. Move the cursor to “Diagnostic variables” and press Enter (F4). Use the ￿ or ￿ key to move up or down the list.

10.2 TROUBLESHOOTING WHEN A FAULT MESSAGE IS SHOWING

 

Fault message

Explanation

See Section

 

 

High input current

input current exceeds 210 uA

10.2.1

 

 

Check sensor zero

sensor current was too high when sensor was zeroed

10.2.2

 

 

Temp error low

temperature less than -15°C

10.2.3

 

 

Temp error high

temperature greater than 130°C

10.2.3

 

 

Sense line open

RTD sense line is open

10.2.4

 

 

Failure factory

instrument needs factory calibration

10.2.5

 

 

Failure eeprom

write verify error has occurred

10.2.5

 

 

 

 

 

 

54

Page 60
Image 60
Emerson 51-DO-03 Section Troubleshooting, Overview, Troubleshooting When a Fault Message is Showing, Measurement Section