Emerson 51-DO-04, 51-DO-03 Alarm setpoints, Output setpoints Simulate tests, Alarm 1 setpoint

Models: 51-DO-04 51-DO-03 DO-04 DO-03

1 70
Download 70 pages 4.9 Kb
Page 30
Image 30

MODEL DO-03/04

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.1 CHANGING ALARM SETPOINTS

1.Before changing alarm setpoints, be sure that alarms are properly configured. See Section 5.6.

Alarm setpoints

Output setpoints

Simulate tests

ExitEnter

Alarm 1 setpoint

Alarm 2 setpoint

Alarm 3 setpoint

ExitEnter

Alarm Low : 0.000 ppm

2.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

3.Press Enter (F4).

4.Move the cursor to the desired alarm and press Enter (F4).

5.A screen like the one shown will appear. The alarm is a low alarm and the setpoint is 0.00 ppm. Press Edit (F4). Use the arrow keys to change the setpoint. Press Save (F4) to store the new value. Press Exit (F1) to return to the screen in step 4. Choose a new alarm.

ExitEdit

24

Page 30
Image 30
Emerson 51-DO-04 Alarm setpoints, Output setpoints Simulate tests, Alarm 1 setpoint, Alarm 2 setpoint, Alarm 3 setpoint