MODEL DO-03/04

SECTION 5.0

 

SOFTWARE CONFIGURATION

TABLE 5-1. Program Settings List

(continued)

 

 

 

 

ITEM

CHOICES

FACTORY SETTINGS

CONFIGURE

 

 

B Outputs (Section 5.5)

 

 

1. Output 1

 

 

a. Measurement

Oxygen or temperature

Oxygen

b. Control

Normal

Normal

2. Output 1 Setup

 

 

a. Current

4-20 mA or 0-20 mA

4-20 mA

b. Dampening

0-299 sec

0 sec

c. Hold mode

Hold last value or go to fixed value

Hold last value

d. Fixed hold value

0-22 mA

21 mA

e. Fault value

0-22 mA

22 mA

3. Output 2

 

 

a. Measurement

Oxygen or temperature

Temperature

b. Control

Normal

Normal

4. Output 2 Setup

See output 1

See output 1

5. Hold feature

Enable, disable, or 20 min timeout

Disable

C. Alarms (Section 5.6)

 

 

1. Alarm 1

 

 

a. Activation method

Oxygen or temperature

Oxygen

b. Control mode

Normal

Normal

2. Alarm 1 setup

 

 

a. Configuration

Low, high, or off

High

b. Hysteresis

 

 

if oxygen (ppm)

0 to 20 ppm

0 ppm

if oxygen (ppb)

0 to 999 ppb

0 ppb

if oxygen (% saturation)

0 to 200%

0%

if temperature

0 to 10°C

0.1°C

c. Delay time

0-99 sec

0 sec

d. Relay fault

none, open, closed

None

3. Alarm 2

 

 

a. Activation method

Oxygen or temperature

Oxygen

b. Control mode

Normal

Normal

4. Alarm 2 setup

 

 

a. Configuration

Low, high, or off

Low

Rest of alarm 2 setup is the same as alarm 1

5.Alarm 3 setup is the same as alarm 1

6.Alarm 4

Alarm

Fault or off

Fault

7. Feed limit timer

 

 

a. Feed limit

Disable, alarm 1, alarm 2, or alarm 3

Disable

b. Timeout value

0 to 10,800 sec

600 sec

8. Interval timer

 

 

a. Select alarm

Disable, alarm 1, alarm 2, or alarm 3

Disable

b. Interval time

0 to 999.9 hr

24.0 hr

c. Repeats

1 to 60

1

d. On time

0 to 2999 sec

120 sec

e. Off time

0 to 2999 sec

1 sec

f. Recovery time

0 to 999 sec

600 sec

22

Continued on the following page

 

Page 28
Image 28
Emerson 51-DO-03, 51-DO-04 instruction manual Choices Factory Settings Configure