SECTION 5.0

 

SOFTWARE CONFIGURATION

5.4 CHOOSING DISPLAY OPTIONS

1.The 54eA analyzer can be used with other oxygen sensors manufactured by Rosemount Analytical as well as with chlorine, monochloramine, and ozone sensors. The user must configure the analyzer to match the sensor being used.

2.The display menu also lets the user customize the third line in the display, change timeout values, choose a lan- guage other than English, and change the display contrast.

3.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

Output setpoints Simulated tests Configure

4. Move the cursor to "Configure" and press Enter (F4).

ExitEnter

Display

Outputs

Alarms

5. With the cursor on "Display", press Enter (F4).

ExitEnter

Measure : Oxygen

Sensor: RMT Standard

Meas units: ppm

ExitEdit

6.The screen at left appears. The default settings shown for “Measure” (Oxygen) and “Sensor” (RMT Standard) are appropriate for the 499ADO sensor. There are no changes to make.

If you wish to change the units, use the ￿ key to move the cursor to "Meas units" and press Edit (F4). Use the ￿ or ￿ keys to scroll through the choices. Press Save (F4) to store the selection.

27