MODEL DO-03/04

SECTION 2.0

 

INSTALLATION

SECTION 2.0

INSTALLATION

2.1 UNPACKING AND INSPECTION

Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analytical immediately.

The Model DO-03/04 consists of the following items.

 

Model or part number

Description

 

 

 

 

 

 

54e-A-01

Model 54eA analyzer

 

 

 

 

 

 

2002577

Mounting bracket for Model 54e, pipe/wall

 

 

 

 

 

 

23544-00

Cable glands kit for Model 54e, qty 5

 

 

 

 

 

 

499ADO-54

Model 499ADO oxygen sensor

 

 

 

 

 

 

24057-00

Smooth membrane retainer for 499ADO sensor

 

 

 

 

 

 

HRMS-02

Handrail clamp assembly (clamp, sensor maintenance kit, and elbow)

 

 

 

 

 

 

24049-00

Air blast washer head

 

 

 

 

 

 

24050-00

Air compressor and enclosure with air hose (115 VAC) option -03 only

 

 

 

 

 

 

24050-01

Air compressor and enclosure with air hose (220 VAC) option -04 only

 

 

 

 

 

 

24051-00

Mounting hardware for air compressor enclosure

 

 

 

 

 

 

24053-00

O-ring kit for sensor washer head (three O-rings and lube)

 

 

 

 

 

The customer must supply a schedule 80 1-1/2 inch PVC pipe boom to immerse the sensor and cleaning head in the aeration basin.

2.2 INSTALLATION

2.2.1 General information

1.Although the equipment is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperatures.

2.Install the equipment in an area where vibrations and electromagnetic and radio frequency interference are minimized or absent.

3.To keep the analyzer watertight, install plugs (provided) in the unused cable openings.

4.Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer.

5.Provide room to allow the sensor and attached piping to be removed from the aeration basin for service. A maintenance bracket (part of PN 24048-00) can be used to rest the PVC pipe boom on the handrail.

Figure 2-1 shows the suggested arrangement of the analyzer, air compressor, and maintenance bracket on the handrail.