MODEL DO-03/04

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.4 CHOOSING DISPLAY OPTIONS (CONTINUED)

Temp units: °C

Output 1: mA

Output 2: mA

ExitEdit

Language: English

Display left: I

Display right: Out 2

ExitEdit

Display contrast: 40

Timeout: On

Timeout value: 10 min

ExitEdit

7.Set the remainder of the display parameters. Use the ￿ and ￿ keys to choose the desired parameter. Then press Edit (F4). Use the ￿ key to move the cursor to the desired selection. Press Save (F4) to store.

 

Temp units

°C or °F

 

 

 

 

 

 

Output 1

mA or % of full scale

 

 

 

 

 

 

Output 2

mA or % of full scale

 

 

 

 

 

 

Language

English, Français, Español, Deutsch, Italiano

 

 

 

 

 

 

Display left

sensor current (I), alarm 1 setpoint (no units),

 

 

 

alarm 3 setpoint (no units), or blank

 

 

 

 

 

 

Display right

sensor current (I), alarm 2 setpoint (no units),

 

 

 

alarm 3 setpoint (no units), output 2, or blank

 

 

 

 

 

 

Display Contrast

00 (lightest)-99 (darkest); the display contrast

 

 

 

changes as the number changes

 

 

 

 

 

 

Timeout

Timeout returns the display from any other screen

 

 

 

to the main display if no key is pressed before the

 

 

Timeout value

 

 

timeout value is exceeded.

 

 

 

 

 

 

 

 

 

Polling address

Identifies analyzer in multi-drop HART applications.

 

 

 

 

 

SECURITY CAUTION

The analyzer uses the timeout value to activate security. Once the analyzer is unlocked by entering a security code, security will not re-activate until a display timeout occurs. If timeout has been turned off, security will never reactivate.