Emerson 51-DO-03, 51-DO-04 instruction manual Choices Factory Settings

Models: 51-DO-04 51-DO-03 DO-04 DO-03

1 70
Download 70 pages 4.9 Kb
Page 29
Image 29

MODEL DO-03/04

SECTION 5.0

 

SOFTWARE CONFIGURATION

TABLE 5-1. Program Settings List

(continued)

 

 

 

 

 

 

ITEM

CHOICES

FACTORY SETTINGS

CONFIGURE (continued)

 

 

D.Temperature compensation (Section 5.7)

 

1. Temperature compensation

Auto or manual

Auto

2.

Manual temperature

-15 to 130°C

25°C

E. Noise Reduction (section 5.8)

 

 

Noise rejection

50 or 60 Hz

60 Hz

F. Barometric Pressure (Section 5.9)

 

 

1.

Barometric pressure

Auto or manual

Auto

2.

Barometric pressure manual

0.4 - 1.2 atm (or equivalent units)

1.00 atm

3.

Barometric pressure units

mm Hg, in Hg, bar, kPa

mm Hg

G. Main sensor calibration (Section 5.10)

 

1.

Stabilize reading

 

 

 

a. oxygen (ppm)

0 to 20 ppm

0.05 ppm

 

b. oxygen (% saturation)

0 to 200%

1%

2.

Stabilize time

0 - 30 sec

10 sec

3.

Sensor zero stabilization value

 

 

4.

Salinity (parts per thousand)

0.0 - 99.9 o/oo

0.0 o/oo

H. Security (Section 5.11)

 

 

1.

Lock all

000-999 (000 disables)

000

2.

Lock program

000-999 (000 disables)

000

3.

Lock configuration

000-999 (000 disables)

000

 

 

 

 

23

Page 29
Image 29
Emerson 51-DO-03, 51-DO-04 instruction manual Choices Factory Settings