SPECIFICATIONS - 499ADO-54 SENSOR
Range: 0 to 20 ppm (mg/L)
Wetted parts: Noryl1 Viton2 EPDM, Teflon3, silicone
Cathode: Gold (not normally wetted)
Accuracy: ±0.2 ppm at 25°C
Repeatability: ±0.05 ppm 25°C
Response time: 25 sec to 63% of final reading at 25°C
Pressure: 0 to 65 psig (0 to 549 kPa abs)
Temperature: 32 to 122°F (0 to 50°C)
Process connection: 1 inch MNPT
Electrolyte volume: 25 mL (approx.)
Cable length (standard): 25 ft (7.6 m)
Cable length (maximum): 300 ft (91 m)
SPECIFICATIONS - 54EA-01 ANALYZER
Enclosure:
(5.7 x 5.7 x 5.2 in.), DIN size panel
Front Panel: Membrane keypad with tactile feedback. Three green LEDs indicate alarm status. Red LED indicates fault condition.
Display:
Power:
115 VAC ± 10%, 50/60 Hz ± 6%, 8 W 230 VAC ± 10%, 50/60 Hz ± 6%, 8 W
RFI/EMI: EN-61326
LVD:
Outputs: Two
Output Accuracy: ± 0.05 mA
Alarms: Relays 1, 2, and 3 are assignable to oxygen or temperature. One relay can also be used as an interval timer to operate the air blast cleaner.
Relay 4 - Sensor/analyzer fault alarm
Each relay has a dedicated LED on the front panel.
Relay Contacts: Relays
Relay 4: Epoxy sealed form C, SPDT
| Resistive | Inductive |
115 Vac | 5.0 Amps | 3.0 Amps |
230 Vac | 5.0 Amps | 1.5 Amps |
Ambient Temperature: 0 to 50°C (32 to 122°F). Analyzer can be operated between
Relative Humidity: 95% (maximum)
Temperature correction for membrane permeability: automatic between 0 and 50°C.
Calibration: automatic air calibration or calibration against a standard instrument
Pressure sensor range: 113 to 862 mm Hg (150 to 1150 mbar)