Hold Button (MA95Z)

SECTION 2—OPERATION

FUNCTION

DESCRIPTION

 

 

PULSE

Pulse mode. The pulse time indicates the time between

 

pulses. For example, 5 minutes is selected. In this

 

example, all of the air cushions in the mattress will

 

deflate for 4 seconds and inflate for 5 minutes (the

 

selected time).

 

 

FOWLER

Fowler Mode. The pressure will increase to prevent the

 

patient from bottoming out.

 

 

MULTI (PULSE + STATIC)

Pulse and Static mode.

 

 

3.Use the Select button to move the pointer to the times or attributes available for the selected function.

4.Use the Firmer or Softer buttons to change the times or attributes.

Hold Button (MA95Z)

1.Press the Hold button (Firm/Soft Buttons) to pause the mattress in the current position. The mattress will remain in this position until the Hold button is pressed again.

Firm/Soft Buttons

1.Select comfort pressure settings by pressing Firm ( Automatic Wireless Fowler) or Soft (Manual Operation of the Fowler) buttons.

Soft Button ‐ Pressing this button reduces the pressure in the mattress.

Firm Button ‐ Pressing this button increases the pressure in the mattress.

NOTE: The patient comfort pressure ranges from Soft (6±4 mmHg) to Firm (32±6 mmHg). The Comfort Control LED displays the patient comfort pressure levels and provides a guide to the caregiver to set approximate comfort pressure level depending on the patient weight. If the patient’s weight to height ratio is above average, increase the pressure setting by approximately 20%. MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from 6 to 32.

Automatic Wireless Fowler

Manual Operation of the Fowler

1.Press the Select button (About Automatic Wireless Mode) to move the pointer on the LCD to Function.

2.Use the Firm and Soft buttons until Fowler displays.

NOTE: When this mode is activated, the control unit increases the pressure in the mattress to prevent the patient from bottoming out.

About Automatic Wireless Mode

When the bed articulates to 45°, the transmitter in the head of the mattress signals the control unit to increase the pressure in the mattress by 80% to prevent the patient from bottoming out.

MA90 Series

16

Part No 1148139