SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

 

 

 

Joystick erratic or does not

Electrical malfunction.

Contact dealer/Invacare for service.

respond as desired.

Controller programmed

Reprogram controller. Contact dealer/Invacare

 

improperly.

for service.

 

 

 

Wheelchair does not

Poor battery terminal con-

Clean terminals.

respond to commands.

nection.

 

Power indicator off - even

Electrical malfunction.

Contact dealer/Invacare.

after recharging.

 

 

 

 

 

NOTE: For additional troubleshooting information and explanation of error codes, refer to the Electronics Manual (P/N 1110532) supplied with each wheelchair.

Checking Battery Charge Level

DON’T

DO

 

 

Don’t perform any installation or maintenance with-

Read and understand this manual and any service infor-

out first reading this manual.

mation that accompanies a battery and charger before

 

operating the wheelchair.

Don’t make it a habit to discharge batteries to the

Recharge as frequently as possible to maintain a high

lowest level.

charge level and extend battery life.

Don’t use randomly chosen batteries/chargers.

Follow recommendations in this manual when selecting a

 

battery or charger.

Don’t put new batteries into service before charging.

Fully charge a new battery before using. Use a carrying

 

strap to remove, move or install a battery.

Don’t tip or tilt batteries.

 

Don’t tap on clamps or terminals with tools.

Push battery clamps onto terminals. Spread clamps wider

 

if necessary.

Don’t mismatch your battery and chargers.

Use ONLY a GEL charger for a GEL or sealed battery

Don’t use randomly chosen batteries or chargers.

and a regular charger for regular batteries.

 

 

P9000XDT