19

Actuator

Cantilever Arm

Spring

FIGURE 4 - REPOSITIONING/REPLACING THE
LOCKING MECHANISM IN THE
CANTILEVER ARM

Angled Portion of

Locking Mechanism

Locking

Mechanism

ARMREST PAD ADJUSTMENT/REPLACEMENT (FIGURE 5)

Armrest Pad Adjustment

1. Remove the phillips screw from the rear of the
armrest pad.
2. Reposition the cantilever slide tube to one (1) of
five (5) positions.
3. Align the mounting hole in the cantilever slide tube
with the desired mounting hole from STEP 2.
4. Install the phillips screw UP through the desired
mounting hole in the armrest tube and the cantile-
ver slide tube mounting hole to secure the arm-
rest pad to the armrest tube.
5. Repeat STEPS 1-4 for opposite armrest pad if
necessary.

Armrest Pad Replacement

1. Remove the phillips screw from the rear of the
armrest pad.
2. Remove the phillips screw from the front of the
armrest pad.
3. Remove the EXISTING armrest pad from the arm-
rest tube.
4. Position the NEW armrest pad onto the armrest
tube and cantilever slide tube.
5. Secure the NEW armrest pad to the armrest tube
and cantilever slide tube with the two (2) phillips
screws.

REAR Phillips Screw

Cantilever Slide Tube

FRONT Phillips Screw

FIGURE 5 - ARMREST PAD ADJUSTMENT/
REPLACEMENT

Armrest

Tube

PROCEDURE 5ARMS
A
R
M
S
REPLACING/REPOSITIONING THECANTILEVER ARM LOCKINGMECHANISM (FIGURE 4)
CAUTION
The locking mechanism is spring loaded. Place
your free hand over the end of the cantilever
arm to prevent the locking mechanism from
springing out of the arm.
1. Flip the cantilever arm up and out of the way.
2. Unscrew the actuator from the locking mechanism.
3. Remove the locking mechanism and spring from
the cantilever arm.

NOTE: Inspect the spring and replace if necessary.

4. Slide the NEW spring and locking mechanism into
the cantilever arm.
5. Position the angled portion of the locking mecha-
nism in one (1) of two (2) ways:
A. ANGLED PORTION FACING UP - Arm will
flip UP.
B. ANGLED PORTION FACING DOWN - Arm
will flip DOWN.
6. Push and hold the locking mechanism into the
cantilever arm.
7. Screw the actuator into the locking mechanism
mounting hole.
8. Flip the cantilever arm up/down to lock in place.
9. Pull down/up on the cantilever arm to ensure it
has locked in place.