Page 11
Image 11

NOTE: If the pressure level is consistently low, check for any leakage at air cells or connecting tubes. If necessary, replace any damaged cells or tubes or contact local qualified dealer for repair.

If the control unit is equipped without audible alarm feature, the Low Pressure Indicator will flash until the low-pressure fault condition is solved. If the control unit is equipped with audible alarm feature, the Low Pressure indicator will flash and the alarm will sound when air pressure is below normal (preset pressure). To silence the alarm, simply press the alarm mute button, and it will be silent for 3 minutes; however, the Low Pressure Indicator still remains flashing. After 3 minutes, alarm buzzer will alert again if the pressure has not yet returned to normal.

4.7 Power Failure Alarm

If the pump unit has a sudden power outage during normal use, the power indicator LED on the control panel will flash " yellow" light, and alarm will sound. To silence the alarm, simply press Power button on the control panel.

When power failure situation occurs, alarm will last until backup battery is low or it can be turned off by manual operation.

Once power failure alarm situation is resolved, it is recommended to turn "OFF" the main power switch located at the lateral side of the pump. If the main power switch is not turned "OFF", once the main power comes back again, the pump will start again at "Standby" status.

5. Cleaning

CAUTION: Ensure the pump unit is disconnected from the main power before cleaning.

5.1 Pump Unit

Wipe with a damp cloth and a mild detergent and keep it away from dust. If other detergent is used, choose one that will have no chemical effects on the surface of the plastic case of the pump unit. All parts should be air dried thoroughly before use.

CAUTION: Do not immerse or soak pump unit. 5.2 Mattress

Do not use Phenolic based products for cleaning.

For suitable and locally available cleaner please ask your distributor.

Wipe down with warm water containing a mild detergent. The cover may also be cleaned using sodium hypochlorite diluted in water. All parts should be air dried thoroughly before use.

CAUTION: Dry the mattress on sunless area after cleaning.