Lumex Syatems 763400A Storage, Maintenance General, Air Filter Replacement, Fuse Replacement

Models: 763400A

1 15
Download 15 pages 11.8 Kb
Page 12
Image 12

6. Storage

1. Pull the CPR tags and make sure all cells are deflated.

2. Lay the mattress out flat and upside down.

3. Protect the air tube coupler by putting it in the inner side of mattress.

4. Roll the mattress from the head end towards the foot end.

5. Foot-end straps can then be stretched around the rolled mattress to prevent unrolling.

7. Maintenance

7.1 General

1. Check main power cord and plug if there is abrasions or excessive wear.

2. Check mattress' cover for signs of wear or damage. Ensure mattress cover and tubes are stubbed together correctly.

3. Check the air tube connections for any damage.

4. Report any problems or malfunctions to local distributors.

7.2 Air Filter Replacement

The air filter should be checked and dusted periodically once per month.

1. Two foam type filters are used. One is located on the back of pump and the other one is at the bottom of the Pump.

2. The filter is reusable and can be washed gently with a mild detergent and water.

3. Suggest to replace air filter regularly if the environment Is dirty. This will help to improve the airflow. Spare filters are available from your local distributor.

7.3Fuse Replacement

1.Disconnect the plug from main power when a fuse is to be replaced.

2.Remove the cover of the fuse holder by means of a small screwdriver.

3.Insert a new fuse of the correct rating in, and replace the cover of the fuse holder back. Check the technical specification for correct fuse rating.

11

Page 12
Image 12
Lumex Syatems 763400A manual Storage, Maintenance General, Air Filter Replacement, Fuse Replacement