MABIS/Duro-Med 04-255-001 manual Memory Clear

Models: 04-255-001

1 28
Download 28 pages 35.37 Kb
Page 15
Image 15

1)Press MENU to enter Feature list

2)Scroll (+/-) to “Preset Pressure”

3)Press MENU button

4)Use (+/-) keys to increase or decrease the pressure in increments of 10. The pressure should be 30mmHg higher than your known Systolic pressure.

5)Press POWER to turn off.

The available range is 150 - 300 mmHg.

MEMORY CLEAR

This feature will only display after measured data has been stored (see Memory Select).

Stored readings can be deleted from each memory bank.

Choose the Memory Bank 1 or 2:

1)Press MENU to enter Feature list

2)Press MENU button for “Memory Select”

3)Scroll (+/-) to select either M1 or M2

4)Press POWER to turn off

NOTE! Pressing Power will allow you to exit “Memory Clear” without deleting any readings.

Memory Clear:

5)Re-enter Feature list by pressing MENU

6)Scroll (+/-) to “Memory Clear” and press MENU

7)Use (+/-) keys to move the cursor in front of the reading to be deleted.

8)Press MENU to delete. The numbering sequence will reflect the change.

Repeat as needed.

9)Press POWER to turn off or to exit.