Accessories for reprocessing with fully demineralised water
Illustration shows
|
|
| |||||||
Water quality plays a key role in instrument | • For a continuous supply of demineralised | • For a continuous supply of demineralised |
| ||||||
reprocessing. Raw water contains salts | water | water |
| ||||||
and minerals which can result in deposits | • Throughput: max. permeate capacity | • Throughput: max. permeate capacity |
| ||||||
on the load and on machine surfaces. Fully | 190 l/h | 190 l/h |
| ||||||
demineralised water is instrumental in pre- | • Reverse osmosis system in stainless- | • |
| ||||||
venting instrument corrosion. In installa- | steel plinth with door and floor tray | neighbouring unit |
| ||||||
tions requiring large volumes of water, re- | Installation of 2 x 5 l supply containers in | • |
| ||||||
verse osmosis installations offer an eco- | plinth | • External dimensions: H 380, W 543, |
| ||||||
nomical alternative to demineralisation | • 2 status LEDs (conductivity/throughflow) | D 302 mm |
| ||||||
cartridges (cf. Chart on Page 41). The use | • Max. yield approx. 50% | • Further features and technical data as per |
| ||||||
of demineralised water can increase the | Salt filtration rate |
| |||||||
economic efficiency of | Water quality approx. |
|
|
|
| ||||
Filtration protects against deposits likely to | (dependent on raw water supply, typically | Optional for |
| ||||||
damage the machine, prevents downtimes | (VEOLIA accessory list) |
| |||||||
and repairs and reduces detergent costs. | Water connection to RO 3⁄4" | • Installation of a |
| ||||||
|
|
|
|
| Permeate outlet 3⁄4" | • Additional connection, e.g. for connection |
| ||
The Miele range now includes the | Concentrate outlet 8 mm water hose | of autoclave or water dispenser tap |
| ||||||
Water supply pressure | • Pressure tank to store demineralised |
| |||||||
osmosis systems from VEOLIA. | Electrical connection 230 V/50 Hz | water |
| ||||||
|
|
|
|
| Rated load 1 kW, fuse rating 10 A | • Connection to water softener |
| ||
|
|
|
|
| indicators |
|
|
|
|
|
|
|
|
| Electricity consumption: 0.6 kW/h* | Indicator lights |
| ||
|
|
|
|
| • Cold water max. 28°C |
|
|
|
|
|
|
|
|
|
| Status: |
|
| |
|
|
|
|
| max. raw water hardness 30°dGH, |
|
|
| |
|
|
|
|
|
| ||||
|
|
|
|
|
| Ready or Stand by |
|
| |
|
|
|
|
| 15°dKH (carbonate hardness) |
|
|
| |
|
|
|
|
|
|
|
|
| |
|
|
|
|
| • Rehingeable door |
| Conductivity and |
|
|
|
|
|
|
| • External dimensions: H 520, W 600, |
|
|
| |
|
|
|
|
|
| throughflow |
|
| |
|
|
|
|
| D 560 mm |
|
|
| |
|
|
|
|
|
| indicators: |
|
| |
|
|
|
|
|
|
| Fault |
| 39 |
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|