Register-Level Programming Chapter 4
Lab-NB User Manual 4-40 © National Instruments Corporation
Programming in Controlled Acquisition Mode
The following programming steps are required for a DAQ operation in controlled acquisition
mode:
1. Select analog input channel, gain, and timebase source for counter A0.
2. Program counter B0 (if necessary).
3. Program counters A0 and A1.
4. Clear the A/D circuitry.
5. Program the sample-interval counter (counter A0).
6. Service the DAQ operation.
Each of these programming steps is explained below.
1. Select analog input channel, gain, and timebase source for counter A0.
The analog input channel and gain are selected by writing to the A/D Configuration Register.
The SCANEN bit must be cleared for DAQ operations on a single channel. See the A/D
Configuration Register bit description earlier in this chapter for gain and analog input channel bit
patterns. If counter B0 is being used as a timebase for counter A0, then the TBSEL bit in the
ADC Configuration Register should be set at this time.
The A/D Configuration Register needs to be written to only when the analog input channel, gain
setting, or other function needs to be changed.
2. Program counter B0 (if necessary).
The following sequence should be used to program counter B0 if it is being used. If counter B0
is not being used, skip to step 3. All writes are 8-bit write operations. All values given are
hexadecimal.
a. Write 36 to the Counter B Mode Register (select mode 3).
b. Write the least significant byte of the timebase count to the Counter B Data Register.
c. Write the most significant byte of the timebase count to the Counter B Data Register. For
example, programming a timebase of 10 µsec requires a timebase count of
10 µsec
0.5 µsec = 20 µsec