AppendixA Specifications
©NationalInstruments Corporation A-5 NI5620 User Manual
Environmental SpecificationsCalibrationinterval ................................ 1 year
Warm-uptime ........................................ 10 minutes
Operating environment
Ambienttemperature ...................... 0 to 50 °C
Humidity.........................................10 to90%, noncondensing
Storage environment
Storagetemperature ........................ –20 to 70 °C
Humidity......................................... 5 to 95%, noncondensing
Maximum altitude ..................................2000 meters
Pollutiondegree .................................... 2
Indoor use only
Power Requirements+3.3VDC (±5%)....................................< 600 mA, 400 mA typical
+5VDC (±5%).......................................< 1.5 A, 1 A typical
+12VDC (±5%) .....................................< 450 mA, 330 mA typical
–12VDC (±5%).....................................< 35 mA, 24 mA typical
Maximum Working VoltageChannelto earth ..................................... 2 V, Installation Category I
SafetyMeets the requirements of the followingstandards for safety for electrical
equipmentfor measurement, control, and laboratoryuse:
EN 61010-1:1993/A2:1995, IEC 61010-1:1990/A2:1995,
UL 3101-1:1993, UL 3111-1:1994, UL 3121:1998,
CAN/CSA C22.2 no. 1010.1:1992/A2:1997 d.