AppendixA Specifications
NI5911 User Manual A-2 ni.com
Sample onboard memory........................4 MB or 16 MB
Memory sample depth
Vertical sensitivity (input ranges)
20 kS/s FlexibleResolu tion 20.5 Bi ts
10 kS/s FlexibleResolu tion 21 B its
*1<n<2
32in oscilloscope mode
Sampling
Frequency Mode
SampleDepth
(4MB)
SampleDepth
(16MB)
100/n*MS/s Oscilloscope 4MS 16MS
12.5MS/s Flexible Resolution 1MS 4MS
5MS/s Flexible Resolution 1MS 4MS
2.5 MS/s Flexible Resolution 1MS 4MS
1MS/s Flexible Resolution 1MS 4MS
500 kS/s Flexible Resolution 1MS 4MS
200 kS/s Flexible Resolution 1MS 4MS
100 kS/s Flexible Resolution 1MS 4MS
50kS/s FlexibleResolution 1MS 4MS
20kS/s FlexibleResolution 1MS 4MS
10kS/s FlexibleResolution 1MS 4MS
*1<n<232 in oscilloscope mode
InputRange NoiseReferred to Input
±10 V 174 dBfs/
±5V 168 dBfs/
±2V 160 dBfs/
±1V 154 dBfs/
SampleRate Mode EffectiveResolution
Hz
Hz
Hz
Hz