AppendixA Specifications
NI5911 User Manual A-4 ni.com
FilteringDynamic RangeNoise (excluding input-referred noise)
Sampling
Frequency
Filter
Mode Bandwidth Ripple
Alias
Attenuation
100/n*MS/s Oscilloscope 100MHz ±3dB N/A
12.5MS/s Flexible
Resolution
3.75MHz ±0.2dB –60dB
5MS/s Flexible
Resolution
2MHz ±0.1dB –70dB
2.5MS/s Flexible
Resolution
1MHz ±0.05dB –80 dB
1MS/s Flexible
Resolution
400kHz ±0.005dB –80dB
500kS/s Flexible
Resolution
200kHz ±0.005dB –80dB
200kS/s Flexible
Resolution
80kHz ±0.005 dB –80dB
100kS/s Flexible
Resolution
40kHz ±0.005 dB –80dB
50kS/s Flexible
Resolution
20kHz ±0.005 dB –80dB
20kS/s Flexible
Resolution
8kHz ±0.005dB –80 dB
10kS/s Flexible
Resolution
4kHz ±0.005dB –80 dB
*1<n<2
32in oscilloscope mode
SamplingFrequency Bandwidth Noise Density Tot al No ise
100/n*MS/s 100MHz –120dBfs/ –43dBfs
12.5MS/s 3.75 MHz –135 dBfs/ –64dBfs
5MS/s 2MHz –150dBfs/ –83dBfs
2.5MS/s 1MHz –155dBfs/ –91dBfs
Hz
Hz
Hz
Hz