AppendixA Specifications
©NationalInstruments Corporation A-7 NI5911 User Manual
Sensitivity...............................................170 steps in full-scale voltage
range
TRIG input range................................... ±5 V (without external
attenuation)
TRIG input impedance........................... 1 M±1% in parallel
with 30 pF ± 15 pF
TRIG input protection ............................±42 V [(DC + peak AC) < 10 kHz,
without external attenuation]
Acquisition Modes
RIS......................................................... 1 GS/s down to 200 MS/s
effectivesample rate, repetitive
signalso nly. Data is interleaved
insoftware.
RIS accuracy .......................................... <0.5 ns
Single-shot............................................. 100 MS/s down to 10 kS/s sample
ratefor t ransient and repetitive
signals
Power Requirements
+5VDC .................................................4 A
+12VDC ................................................ 100 mA
12VDC ............................................... 100 mA
Physical
Dimensions.............................................33.8 by 9.9 cm
(13.3by3.9in)
I/O connectors
Analoginput CH0........................... BNC female
Digitalt riggers................................ SMB female, 9-pin mini DIN