Chapter 4 Connecting Signals
NI 6115/6120 User Manual 4-6 ni.com
Table 4-4. Analog I/O Summary for the NI 6120Signal Name
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA atV)
Rise
Time
(ns) Bias
ACH<0..3>+ AI 100 GΩ
to GND
±42 V to
GND
— — — ±300 pA
ACH<0..3>– AI 100 GΩ
to GND
±42 V to
GND
— — — ±200 pA
Differential Pair
ACH<0..3>+ to
ACH<0..3>–
AI 1 MΩ in parallel
with 100 pF1 or
10k in parallel
with 40 pF2
— — — — ±300 pA
DAC0OUT AO 50ΩShort-circuit
to ground
5 at 10 5 at –10 — —
DAC1OUT AO 50ΩShort-circuit
to ground
5 at 10 5 at –10 — —
1 Applies to range ≤ ±10 V, impedance refers to ACH<0..3>–.
2 Applies to range > ±10 V, impedance refers to ACH<0..3>–.
Table 4-5. Digital I/O Signal SummarySignal Name
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA atV)
Rise
Time
(ns) Bias
VCC DO 0.1ΩShort-circuit
to ground
1 A — — —
DIO<0..7> DIO — VCC +0.5 13 at
(VCC–0.4)
24 at 0.4 1.1 50kΩ
pu
SCANCLK DO — — 3.5 at
(VCC–0.4)
5 at 0.4 1.5 50kΩ
pu
EXTSTROBE* DO — — 3.5 at
(VCC–0.4)
5 at 0.4 1.5 50kΩ
pu
PFI0/TRIG1 AI
DIO
10kΩ±35
VCC +0.5
3.5 at
(VCC–0.4)
5 at 0.4 1.5 9kΩ
pu and
10kΩ
pd
PFI1/TRIG2 DIO — VCC +0.5 3.5 at
(VCC–0.4)
5 at 0.4 1.5 50kΩ
pu
PFI2/CONVERT* DIO — VCC +0.5 3.5 at
(VCC–0.4)
5 at 0.4 1.5 50kΩ
pu