© National Instruments Corporation 59 NI cDAQ-9172 User Guide and Specifications
You can route Counter 1 Internal Output to Counter 0 HW Arm. You can
also route Counter 0 Internal Output to Counter 1 HW Arm.
Some of these options may not be available in some driver software.
Counter n Internal Output and Counter n TC SignalsCounter n TC is an internal signal that asserts when the counter value is 0.
The Counter n Internal Output signal changes in response to Counter n TC.
The two software-selectable output options are pulse output on TC and
toggle output on TC. The output polarity is software-selectable for both
options.
Routing Counter n Internal Output to an Output Terminal
You can route Counter n Internal Output to any output PFI terminal.
Frequency Output SignalThe Frequency Output (FREQ OUT) signal is the output of the frequency
output generator.
Routing Frequency Output to a Terminal
You can route Frequency Output to any output PFI terminal. The FREQ
OUT signal can also route to DO Sample Clock and DI Sample Clock.
Default Counter/Timer RoutingCounter/timer signals are available to correlated digital I/O CSeries
modules in slots 5 and/or 6. To determine the signal routing options for
modules installed in your system, refer to the Device Routes tab in MAX.
Counter TriggeringCounters support three different triggering actions—arm start, start, and
pause.
Arm Start TriggerTo begin any counter input or output function, you must first enable, or
arm, the counter. Software can arm a counter or configure counters to be
armed on a hardware signal. Software calls this hardware signal the Arm
Start Trigger. Internally, software routes the Arm Start Trigger to the
Counter n HW Arm input of the counter.