© National Instruments Corporation 7 NI cDAQ-9172 User Guide and Specifications
Setting Up the NI cDAQ-9172Complete the following steps to prepare the NI cDAQ-9172 chassis for use:
1. Before connecting the hardware, install NI-DAQmx software,
VILogger, and the NI-DAQ Device Documentation Browser. Refer to
the DAQ Getting Started Guide for more information about software
installation.
Note The NI-DAQmx software is included on the CD shipped with your kit and is
available for download at ni.com/support. After you install it, the NI-DAQ Device
Documentation Browser is available from Start»Programs»National Instruments»
NI-DAQ»Browse Device Documentation. The DAQ Getting Started Guide is available
after installation from Start»Programs»National Instruments»NI-DAQ»DAQ Getting
Started Guide.
2. If you are not using any mounting accessories, attach the provided
rubber standoffs to the back of the NI cDAQ-9172 chassis.
3. Make sure the NI cDAQ-9172 chassis power switch is turned off.
4. Attach a ring lug to a 14 AWG (1.6 mm) wire. Connect the ring lug to
the ground terminal on the side of the chassis using the ground screw.
Attach the other end of the wire to the system safety ground.
Figure 6. Ring Lug Attached to Ground Terminal
Note Additionally, attach a wire with a ring lug to all other C Series I/O module cable
shields. You must connect this wire to the ground terminal of the chassis using the ground
screw.
1 Attached to System Ground
1