Chapter 3 Signal Connections
© National Instruments Corporation 3-21 PCI-1200 User Manual
Figure 3-11. Pretrigger DAQ Timing
For interval scanning data acquisition, counterB1 determines the scan
interval. Instead of using counterB1, you can externally time the scan
interval through OUTB1. If you externally time the sample interval, we
recommend that you also externally time the scan interval.
Figure3-12 shows an example of an interval scanning DAQ operation. The
scan interval and the sample interval are being timed externally through
OUTB1 and EXTCONV*. Channels 1 and 0 of the input multiplexers are
scanned once during each scan interval. The first rising edge of
EXTCONV* must occur a minimum of 50ns after the rising edge on
OUTB1. The first rising edge of EXTCONV* after the rising edge of
OUTB1 enables an internal GATE signal that allows conversions to occur.
The first conversion then occurs on the following falling edge of
EXTCONV*. The GATE signal disables conversions for the rest of the scan
interval after the desired channels have been scanned. Refer to the Interval
Scanning Acquisition Mode section in Chapter4, Theory of Operation, for
more information on interval scanning.
t
w
50 ns minimum
EXTTRIG
EXTCONV*
CONVERT
V
IH
V
IL
t
w
t
w