Chapter 5 Using the SCXI-1125
© National Instruments Corporation 5-23 SCXI-1125 User Manual
Perform Offset Null Compensation Using Traditional NI-DAQ (Legacy) C APIAfter configuring the system settings and module properties, you
can perform an offset null compensation programmatically using
SCXI_Calibrate. SCXI_Calibrate takes measurements and adjusts the
coarse and fine offset null potentiometers to minimize or eliminate any
electrical offset for a channel. Repeat this process for each channel by
calling the SCXI_Calibrate function in a loop. Use the resulting
imbalance in your application as a software correction factor by
determining the residual voltage from the imbalance, and subtracting this
residual offset from each future measurement. For more information
regarding the operation of SCXI_Calibrate, refer to the Traditional
NI-DAQ (Legacy) Function Reference Help installed by default in Start»
Programs»National Instruments»NI-DAQ.
Perform Acquisition Using Traditional NI-DAQ (Legacy) C APIThere are several NI-DAQ functions you can use to take measurements.
Usually in SCXI the preference is to take multiple samples from multiple
channels using the SCAN_Op function. SCAN_Op performs a synchronous,
multiple-channel scanned data acquisition operation. SCAN_Op does not
return until Traditional NI-DAQ (Legacy) acquires all the data or an
acquisition error occurs (MIO, AI, and DSA devices only). For this reason,
it is sometimes useful to use SCAN_Op in conjunction with the function
Timeout_Config, which establishes a timeout limit synchronous
functions to ensure that these functions eventually return control to your
application. After acquiring data using SCAN_Op, the resultant data is not
organized by channel, so you should demultiplex the data using
SCAN_Demux. SCAN_Demux rearranges, or demultiplexes, data acquired by
a SCAN_Op into row-major order, meaning each row of the array holding
Table 5-8. NI-DAQ Functions Used to Configure SCXI-1125
Channel
Setting NI-DAQ Function to Use
Significant
Parameters
Possible Parameters
Valu es
Gain SCXI_Set_Gain f64 gain
(gain setting)
1, 2, 5, 10, 20, 50, 100,
200, 250, 500, 1000,
2000
Bandwidth SCXI_Configure_Filter f64 freq
(filter cutoff
frequency if
filterMode = 1)
4.0, 10,000.0 Hz