© National Instruments Corporation D-1 SCXI-1125 User Manual
D

Common Questions

This appendix lists common questions related to the use of the SCXI-1125.
The SCXI-1125 is backward compatible with the SCXI-1120, but what
are the major differences between the SCXI-1120 and the SCXI-1125?
Table D-1 compares the major specifications and features of the two
modules. Other specifications and features of the SCXI-1125 are the same
or very similar to the SCXI-1120.
Table D-1. Comparison of the SCXI-1125 with the SCXI-1120
Feature SCXI-1120 SCXI-1125
Analog input 8 8
Input range ±5 V, ±250 V with SCXI-1327 ±5 V, ±300 V with SCXI-1327,
±1000 VDC with TBX-1316
Isolation 250 Vrms 300 Vrms
Gains 1, 2, 5, 10, 20, 50, 100, 200, 250, 500,
1000, and 2000 jumper selectable
1, 2, 5, 10, 20, 50, 100, 200, 250, 500,
1000, and 2000 software selectable
Filters 4 Hz or 10 kHz jumper selectable 4 Hz or 10 kHz software selectable
Autozero Not supported Software configurable and scannable
Calibration Manually rotateable potentiometers
for one-point offset calibration
Software internal one-point offset
calibration, software external
two-point offset and gain calibration,
and onboard calibration constant
storage in EEPROM
Scanning 333 kS/s with consecutive channels 333 kS/s with nonconsecutive and
repeating channels (random scanning)
CJC scanning M TEMP (non-scannable) or
D TEMP (direct channel)
M TEMP, CJ TEMP (scannable)
Offset error ±6 µV ± 3 mV/gain 0.2 mV/gain typical
Gain error ±0.2% typ, ±0.6% max ±0.03% typ, ±0.08% max