Cold Junction adjustment:
After setting Heat/Cool mode, the next parameter is the cold junction reference temperature adjustment. The display will first show 'COLd JN' , and then the cold junction temperature will be indicated. Use ‘^DIGv’ key to adjust until the display reads the proper temperature. Once the correct temperature is displayed, push ‘SETUP’ key to enter that setting and go to temperature Calibration mode.
NOTE 1: The unit comes
NOTE 2:
NOTE 3: No
Thermocouple Calibration:
For
Thermocouple Calibration Procedure
Note: Make sure the unit is reading correct
If incorrect, adjust as described in the " Cold Junction adjustment" section.
For thermocouple calibration (type J T/C), following steps should be performed. Note that only one type of thermocouple need to be calibrated (i.e. J,K,T or E). For example, if the calibration is done for a type K thermocouple, types J, T, and E are automatically calibrated.
1.Connect a thermocouple calibration source to the signal input connector.
2.Dial in 725 degrees centigrade (Note: unit must be programmed for displaying in centigrade).
3.Adjust gain pot on the back of the instrument (ref. Fig. 7, Pg. 21) until the display reads '725'.
4.Short the input connector with a wire or shorting bar (make sure the source is removed so as not to damage it
5. Push ‘SETPT’ key. The display will read 'U 00002'
6.Adjust offset pot on back of instrument (ref. Fig. 7, Pg. 21) until the display reads '0000'.
7.Push SETUP key once
8.Remove the shorting bar from the input and connect the thermocouple calibrator again.
9.Repeat steps 2 through 7 till the unit reads proper temperature.