Page 20
DRX-ACV SPECIFICATIONSINPUTRANGES:
400 mV, 4V, 40 Volt, 400 Volt
FREQUENCYRANGE:
30 Hz to 1 KHz
INPUTIMPEDANCE:
2.1 Meg for all ranges.
ISOLATION:
Dielectric strength to 1000 Vrms transient per
1 minute test based on EN 61010 for 50 Vdc or Vrms
working voltage.
Three way Isolation:
• Power to input
• Power to analog output/communication
• Input to analog output/communication
INPUTOVER-VOLTAGE PROTECTION:
10% Above full scale continuously.
100% Above full scale for 10 seconds.
ANALOG TO DIGITALTECHNIQUE:
Dual slope
READ RATE:
3 readings/second
ACCURACYAT 25 °C:
400 mV, 4V, 40Vand 400Vranges
49 Hz to 500 Hz ±0.2% of FS
30 Hz to 1KHz ±0.2% of FS ±10 counts
TEMPERATURE STABILITY:
400 mVand 40 Volt range 150 ppm/°C typical
4Vand 400 Volt range 100 ppm/°C typical
STEPRESPONSE FOR RS485 OUTPUT:
2 seconds to 99% of the final value
(Filter time constant = 64)
RESPONSE TIME:
To verify the response time, check the carriage return
<CR>, it will be sent at the end of the response. You
can send another command after you receive the
<CR>.
i.e. send: *01X01
response: 01X01<DATA><CR>
Note: 01 is address.
WARM UPTO RATED ACCURACY:
30 minutes
INPUTPOWER:
10 to 32 Volt DC
POWER CONSUMPTION:
2.4 watts (100 mAat 24 Vdc)
OPERATING AMBIENT:
-5 to +55 °C
STORAGE TEMPERATURE:
-40 to +85 °C
RELATIVE HUMIDITY:
90% at 40 °C non-condensing
SIG HI
N/C
N/C
SIG LO
+
+-
N/C
SIGNAL INPUT
D
C
P
W
R
1 2