Calibration- Thermocouple Type K, J and

T

and E

Equipment List

4 ½ digit voltmeter

Voltage Calibrator:

Resolution, 1uV; accuracy ±0.01% (±2uV)

Ice Bath, 0.00 °C ± 0.05°C

Jumper lead terminated in mini-clips

For optimum results, the calibration should be performed with an ambient temperature from 20° 24°C (68° to 75°F). Refer to the figure on the previous page for the locations of all testpoints and adjustments.

a.Turn the indicator ON by pressing the °C key.

b.For types K, J, E and T units only, switch the indicator to the 0.1° range.