Calibration- Thermocouple Type K, J and
T
and E
Equipment List
4 ½ digit voltmeter
Voltage Calibrator:
Resolution, 1uV; accuracy ±0.01% (±2uV)
Ice Bath, 0.00 °C ± 0.05°C
Jumper lead terminated in mini-clips
For optimum results, the calibration should be
performed with an ambient temperature from 20°
24°C (68° to 75°F). Refer to the figure on the
previous page for the locations of all testpoints
and adjustments.
a. Turn the indicator ON by pressing the °C key.
b. For types K, J, E and T units only, switch
the indicator to the 0.1° range.