Calibration- Thermocouple Type K, J and
T
and E
Equipment List
∙4 ½ digit voltmeter
∙Voltage Calibrator:
Resolution, 1uV; accuracy ±0.01% (±2uV)
∙Ice Bath, 0.00 °C ± 0.05°C
∙Jumper lead terminated in
For optimum results, the calibration should be performed with an ambient temperature from 20° 24°C (68° to 75°F). Refer to the figure on the previous page for the locations of all testpoints and adjustments.
a.Turn the indicator ON by pressing the °C key.
b.For types K, J, E and T units only, switch the indicator to the 0.1° range.
19