NOTE: If you want to cancel the calibration mode press OFF button in front of the unit.
For T/C type | K set simulator @ 1320 °C |
| J set simulator @ 960 °C |
| T set simulator @ 350 °C |
| E set simulator @ 950 °C |
Press the switch one more time and wait for a few seconds until display switches to 0.0 set simulator to 0°C. Press switch one more time. Unit should
Step 4. Now, turn the unit on from front panel. Make sure the unit in degree C mode. Verify that the unit is within specification throughout its range.
Step 5. If unit does not meet specification as 0°C then repeat calibration procedure,
Instead of setting simulator as 0°C, set it at the temperature indicated by the unit in measurement mode as 0°C. For example in measure mode, unit may display .2°C with simulator as 0°C. During