Zero Calibration

a.Turn the indicator on by pressing the °C key.

b.Set the indicator to the 0.1° range.

c.Connect the ice bath probe to the indicator Input jack.

d.Place the probe in a stable ice bath (0.00C

±0.05°C and adjust the ZERO pot (Rb) located beneath the plug on the rear case for an indicator display of 0.0°C.

e.Replace the rubber plug.

Calibration is complete.