GENERAL METER SPECIFICATIONS
1. DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digits
2. POWER: Input voltage range is +9 to +28 VDC with short circuit and input polarity protection. Must use a
MODEL | DISPLAY COLOR | INPUT CURRENT | INPUT CURRENT | |
@ 9 VDC WITHOUT | @ 9 VDC WITH | |||
NO. | ||||
| ||||
|
| |||
|
|
|
| |
10 mA | 40 mA | |||
Red (max intensity) | 85 mA | 115 mA | ||
Green (max intensity) | 95 mA | 125 mA |
3. INPUT RANGES: Jumper Selectable
D.C. Currents: 200 ∝A, 2 mA, 20 mA, or 200 mA
4. SIGNAL INPUTS:INPUT | ACCURACY | INPUT | MAX |
| TEMP. | |
@23 °C, less | INPUT | RESOLUTION | ||||
RANGE | IMPEDANCE | COEFFICIENT | ||||
| than 85% RH |
| SIGNAL |
|
| |
200 ∝A | 0.1% of span | 1.111 KΩ | 15 mA | 10 nA | 70 ppm / °C | |
2 mA | 0.1% of span | 111 Ω | 50 mA | .1 µA | 70 ppm / °C | |
20 mA | 0.1% of span | 11 Ω | 150 mA | 1 µA | 70 ppm / °C | |
200 mA | 0.1% of span | 1 Ω | 500 mA | 10 µA | 70 ppm / °C |
9 to 28 VDC power circuit is not isolated from the signal circuit. Input Overrange Indication: “OLOL”.
Input Underrange Indication: “ULUL”.Display Overrange/Underrange Indication:8.USER INPUT (USR): Programmable input. Connect terminal to common (USR COMM) to activate function. Internal 10KΩ
Threshold Levels: VIL = 1.0 V max; VIH = 2.4 V min; VMAX = 28 VDC
Response Time: 5 msec typ.; 50 msec debounce (activation and release)
9.MEMORY: Nonvolatile E2PROM memory retains all programming parameters and max/min values when power is removed.
10. ENVIRONMENTAL CONDITIONS:Operating Temperature Range forOperating Temperature Range for
| INTENSITY LEVEL | TEMPERATURE |
Red Display | 1 & 2 | |
| 3 | |
| 4 | |
| 5 | |
Green Display | 1 & 2 | |
| 3 | |
| 4 | |
| 5 |
Operating and Storage Humidity: 0 to 85% max. relative humidity (non- condensing)
Vibration According to IEC
Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directions
Altitude: Up to 2000 meters
11.CONNECTIONS: Wire clamping screw terminals Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 5 inch-lbs (0.565 N-m) max.
12.CONSTRUCTION: This unit is rated for NEMA 4X/IP65 requirements for indoor use. Installation Category I, Pollution Degree 2. High impact plastic case with clear viewing window. Panel gasket and mounting clip included.
13.CERTIFICATIONS AND COMPLIANCES:SAFETYUL Recognized Component, File #E313607,
UL Listed, File # E313547, UL508, CSA C22.2 No.
IEC
IP65 Enclosure rating (Face only), IEC 529
Type 4X Indoor Enclosure rating (Face only), UL50
ELECTROMAGNETIC COMPATIBILITYEmissions and Immunity to EN 61326: Electrical Equipment for Measurement, Control and Laboratory use.
Immunity to Industrial Locations: |
| |
Electrostatic discharge | EN | Criterion A |
|
| 4 kV contact discharge |
|
| 8 kV air discharge |
Electromagnetic RF fields | EN | Criterion A |
|
| 10 V/m |
Fast transients (burst) | EN | Criterion A |
|
| 2 kV power |
|
| 1 kV signal |
Surge | EN | Criterion A |
|
| 1 kV |
|
| 2 kV |
RF conducted interference | EN | Criterion A |
|
| 3 V/rms |
Power frequency magnetic fields EN | Criterion A | |
Emissions: |
| 30 A/m |
|
| |
Emissions | EN 55011 | Class A |
Notes:
1.Criterion A: Normal operation within specified limits.
14.WEIGHT: 3.2 oz (100 g)
PART NUMBER INFORMATION
DESCRIPTION | PART NUMBER |
|
|
DC Current Meter with reflective display | |
|
|
DC Current Meter with backlight display | |
|
|
Single Relay Output Card | |
|
|
Dual Sinking Open Collector Output card | |
RS485 Serial Communications Card | |
|
|
RS232 Serial Communications Card | |
|
|
|
|
RS232 Programming Cable | |
|
|
RS485 Programming Cable | |
|
|
PC Configuration Software for Windows 98, ME, 2000, XP * |