5.Handling Errors and Problems
5. Handling Errors and Problems
5.1Error Messages
Error Display | Cause | Remedy |
|
|
|
|
| Press air release button and |
| Cuff is under inflated. | restart the measurement with a |
| higher inflation level. Carefully | |
|
| |
|
| read steps under section 3.3. |
|
|
|
|
| Repeat measurement. Remain |
| Movement during | still and do not talk during |
| measurement | measurement. |
|
| Refer to section 3.3. |
|
|
|
| Air tube disconnected. | Insert the air tube securely. |
| Refer to section 3.2. | |
|
| |
|
|
|
| Arm cuff not applied | Apply the arm cuff correctly. |
| correctly. | Refer to section 3.2. |
|
|
|
| Clothing is interfering | Remove any clothing interfering |
| with the arm cuff. | |
| with the arm cuff. | |
| Refer to section 3.2. | |
|
| |
|
|
|
| Air is leaking from the | Replace cuff with new one. |
| arm cuff. | Refer to Chapter 7. |
|
|
|
|
| Release your hand from the air |
| The arm cuff was inflated | inflation bulb before the pressure |
| above 299 mmHg. | reaches 299 mmHg. |
|
| Refer to section 3.3. |
| Replace all four “AAA” batteries |
Battery power is low. | with new ones. |
| Refer to section 2.1. |
Blinks or appears
continuously
EN |
27