5.Handling Errors and Problems

5. Handling Errors and Problems

5.1Error Messages

Error Display

Cause

Remedy

 

 

 

 

 

Press air release button and

 

Cuff is under inflated.

restart the measurement with a

 

higher inflation level. Carefully

 

 

 

 

read steps under section 3.3.

 

 

 

 

 

Repeat measurement. Remain

 

Movement during

still and do not talk during

 

measurement

measurement.

 

 

Refer to section 3.3.

 

 

 

 

Air tube disconnected.

Insert the air tube securely.

 

Refer to section 3.2.

 

 

 

 

 

 

Arm cuff not applied

Apply the arm cuff correctly.

 

correctly.

Refer to section 3.2.

 

 

 

 

Clothing is interfering

Remove any clothing interfering

 

with the arm cuff.

 

with the arm cuff.

 

Refer to section 3.2.

 

 

 

 

 

 

Air is leaking from the

Replace cuff with new one.

 

arm cuff.

Refer to Chapter 7.

 

 

 

 

 

Release your hand from the air

 

The arm cuff was inflated

inflation bulb before the pressure

 

above 299 mmHg.

reaches 299 mmHg.

 

 

Refer to section 3.3.

 

Replace all four “AAA” batteries

Battery power is low.

with new ones.

 

Refer to section 2.1.

Blinks or appears

continuously

EN

27

Page 27
Image 27