Omron M24/7 Recommended separation distance, Separation distance in function of the frequency

Models: M24/7

1 28
Download 28 pages 12.62 Kb
Page 25
Image 25

Recommended separation distance

OMRON M24/7 is intended to be used in electromagnetic environment with controlled RF disturbances. The purchaser or user of M24/7may help to reduce electromagnetic disturbances by de ning the separation distance between the transportable or mobile RF telecommunication equipment (transmitters) and OMRON M24/7, depending on the highest output power of the telecommunication equipment.

 

Separation distance in function of the frequency of the

 

transmitter [m]

 

 

 

The highest

150KHz –

80MHz –

 

800MHz – 2.5GHz

output

80MHz

800MHz

 

d=[7/E1] P

power of the

d=[3.5/V1]

d=[3.5/E

] P

 

transmitter [W]

 

1

 

 

 

 

 

 

0.01

Not applicable

0.12

 

0.23

0.1

Not applicable

0.38

 

0.73

1

Not applicable

1.2

 

2.3

10

Not applicable

3.8

 

7.3

100

Not applicable

12

 

23

If this table does not contain the highest output power of the transmitter, the d separation distance [m] can be calculated by the formula, depending on the frequency of the transmitter, where P is the rated highest output power of the transmitter [W].

1.note: in case of frequency 80MHz or 800 MHz, the formula for the higher range is applicable.

2.note: These are guidelines. Actual conditions may vary.

25

Page 25
Image 25
Omron M24/7 Recommended separation distance, Separation distance in function of the frequency, =3.5/E Transmitter W