Download 28 pages 12.62 Kb
Page 22
Image 22

4.3. EMC Information

Medical electrical equipment should be used with precautions according to EMC, and must be installed according to the EMC notices disclosed in this manual as mobile RF transceivers could adversely affect it.

Directive and declaration of manufacturer – Electromagnetic Emission

The OMRON M24/7 is suitable for use in the speci ed electromagnetic environment. The purchaser or user of the OMRON M24/7 should assure that it is used in an electromagnetic environment as described below

Emission test

Compliance

Electromagnetic Environment

Radiated and conducted

Group 1

OMRON M24/7 uses RF energy only for its

RF emission

 

internal function. Therefore, the emission

CISPR 11

 

is very low and not likely to cause any

 

interference in nearby electronic equipment.

 

 

Radiated and conducted

Class B

OMRON M24/7 is suitable for use in domestic

RF emission

 

establishments and in establishments directly

CISPR 11

 

connected to the low voltage power supply

 

network which supplies buildings used for

 

 

 

 

domestic purposes.

Harmonic emission

Not applicable

---

IEC61000-3-2

 

 

Voltage uctuations /

Not applicable

---

Flickers

 

 

IEC61000-3-3

 

 