Directive and declaration of manufacturer – Electromagnetic immunity

OMRON M24/7 is suitable for use in the speci ed electromagnetic environment. The purchaser or user of OMRON M24/7 should assure that it is used in an electromagnetic environment as described below.

Immunity test

IEC60601-1-2

Compliance

Electromagnetic

 

test level

level

environment

Electrostatic

±6KV contact

±8KV air

Floors are wood, concrete or

discharge (ESD)

±8KV air

 

ceramic tile, or oors are covered

IEC 61000-4-2

 

with synthetic material and the

 

 

 

 

relative humidity is at least 30

 

 

 

 

 

 

percent.

Electrical fast

±2KV for power

Not applicable

Mains power quality is that of a

transient/burst

supply lines

 

typical commercial and/or hospital

IEC 61000-4-4

±1KV for input/

 

environment.

 

 

 

output lines

 

 

Surge

±1KV differential

Not applicable

Mains power quality is that of a

IEC 61000-4-5

mode

 

typical commercial and/or hospital

±2KV common

 

environment.

 

 

 

 

 

 

mode

 

 

Voltage dips, short

<5% UT (>95%

Not applicable

Mains power quality is that of

interruptions and

dip) for 0.5 cycle;

 

a typical commercial and/or

voltage variations

40% UT (60% dip)

 

hospital environment. If the user of

on power supply

 

OMRON M24/7 requires CLINICAL

for 5 cycles;

 

input lines

 

UTILITY during power mains

70% UT (30% dip)

 

IEC 61000-4-11

 

interruptions, it is recommended

for 25 cycles;

 

that parts of the OMRON M24/7

 

 

 

<5% UT (>95%

 

system where applicable be

 

 

powered from an uninterruptible

 

dip) for 5 sec.

 

 

 

power supply.

 

 

 

Power frequency

3A/m

3A/m

Power frequency magnetic elds

(50/60Hz) magnetic

 

 

are at levels characteristic of

eld

 

 

a typical location in a typical

IEC 61000-4-8

 

 

commercial and/or hospital

 

 

environment.

 

 

 

Note: UT is the nominal voltage of mains.