9. Technical data
Technical data for the OMRON Comp Air compressor with the OMRON V.V.T. (Virtual Valve Technology) Nebuliser kit:
Particle Size:
Medication Tank
Capacity:
Appropriate Medication
Quantities:
Sound:
Nebulisation Rate:
Aerosol Output:
Aerosol Output Rate:
*MMAD **approximately 3µm 7 ml maximum
2 ml minimum - 7 ml maximum Noise level (at 1 m distance) 53 db
0.35ml/min (by weight loss) **0.3 ml (2ml, 1%NaF) **0.05 ml/min (2ml, 1%NaF)
Notes:
•Performance may vary with drugs such as suspensions or high viscosity. See drug supplier’s data sheet for further details.
•The data above is referenced by Omron Healthcare.
•See web site of Omron Healthcare Europe to update technical information.
Result of cascade impactor** measurements for particle size with OMRON
Cumulative % particle mass of sodium fluoride under size
| 100 |
|
|
|
| 90 |
|
|
|
Undersize% | 80 |
|
|
|
70 |
|
|
| |
60 |
|
|
| |
50 |
|
|
| |
Cumulative |
|
|
| |
40 |
|
|
| |
30 |
|
| Individual tests | |
|
|
| ||
|
|
|
| |
| 20 |
|
| Mean |
| 10 |
|
| MMAD: 3µm |
|
|
|
| |
| 0 |
|
|
|
| 0.1 | 1 | 10 | 100 |
Particle Size Dp (µm)
*MMAD = Mass Median Aerodynamic Diameter
**Independently measured at SolAero Ltd., Canada, Dr. John Dennis, according to EN
EN
27