9. Technical data

Technical data for the OMRON Comp Air compressor with the OMRON V.V.T. (Virtual Valve Technology) Nebuliser kit:

Particle Size:

Medication Tank

Capacity:

Appropriate Medication

Quantities:

Sound:

Nebulisation Rate:

Aerosol Output:

Aerosol Output Rate:

*MMAD **approximately 3µm 7 ml maximum

2 ml minimum - 7 ml maximum Noise level (at 1 m distance) 53 db

0.35ml/min (by weight loss) **0.3 ml (2ml, 1%NaF) **0.05 ml/min (2ml, 1%NaF)

Notes:

Performance may vary with drugs such as suspensions or high viscosity. See drug supplier’s data sheet for further details.

The data above is referenced by Omron Healthcare.

See web site of Omron Healthcare Europe to update technical information.

Result of cascade impactor** measurements for particle size with OMRON NE-C30-E and OMRON V.V.T. Nebuliser kit

Cumulative % particle mass of sodium fluoride under size

 

100

 

 

 

 

90

 

 

 

Undersize%

80

 

 

 

70

 

 

 

60

 

 

 

50

 

 

 

Cumulative

 

 

 

40

 

 

 

30

 

 

Individual tests

 

 

 

 

 

 

 

 

20

 

 

Mean

 

10

 

 

MMAD: 3µm

 

 

 

 

 

0

 

 

 

 

0.1

1

10

100

Particle Size Dp (µm)

*MMAD = Mass Median Aerodynamic Diameter

**Independently measured at SolAero Ltd., Canada, Dr. John Dennis, according to EN 13544-1

EN

27

Page 27
Image 27
Omron NE-C30-E instruction manual Mmad = Mass Median Aerodynamic Diameter